Your browser doesn't support javascript.
loading
Avaliação do desempenho elétrico de um cabo-eletrodo bipolar endocárdico de fixação ativa estimulação atrial direita / Evaluation of the electrical performance of a bipolar active fixation endocardial lead in right atrial pacing
Dutra, Luiz Ricardo Pescatori.
Affiliation
  • Dutra, Luiz Ricardo Pescatori; Santa Casa de Misericórdia. Araraquara. BR
RELAMPA, Rev. Lat.-Am. Marcapasso Arritm ; 27(2): 90-93, abr.-jun.2014.
Article in Pt | LILACS | ID: lil-727488
Responsible library: BR44.1
RESUMO

Introdução:

A evolução da tecnologia envolvida no desenvolvimento e na construção dos cabos-eletrodosvisa a melhorar seu desempenho elétrico na prática clínica da estimulação cardíaca artificial, otimizando o usodos sistemas e tornando-os mais duradouros, sem perder a eficácia. Neste estudo, objetivou-se avaliar a efetividadee a segurança de um cabo-eletrodo bipolar endocárdico de fixação ativa St. Jude, modelo Tendril 2088TC®.

Método:

Foi realizada uma análise prospectiva de 43 pacientes submetidos a implante atrial direito do modelo emquestão, dos quais 26 (60%) eram do sexo feminino. Os valores de limiar de estimulação, sensibilidade da ondaP e impedância foram compilados em quatro momentos da evolução clínica de cada paciente no implante, naalta hospitalar, 30 dias após o implante e com três meses de evolução. Ao final do período de estudo, 41 pacientescompletaram o seguimento.

Resultados:

Após análise estatística, os resultados demonstraram desempenhoelétrico favorável, com valores crônicos estáveis.

Conclusão:

Conclui-se que o referido cabo-eletrodo é eficiente eseguro na rotina do serviço de marcapasso.
ABSTRACT

Background:

The evolution of the technology used for the development and construction of leads,is aimed at improving their electrical performance in the clinical practice of artificial cardiac pacing, optimizingsystems for a longer duration without losing their efficacy. The objective of this study is to evaluate the efficacyand safety of the St. Jude’s bipolar active fixation endocardial lead, model Tendril 2088TC®.

Method:

Prospectivestudy of 43 patients undergoing right atrial implantation using the model above, of which 26 (60%) were female.The threshold values for stimulation, P-wave sensitivity and impedance were collected at 4 timepoints duringthe clinical follow-up of each patient at the time of implantation, at hospital discharge, 30 days and 3 monthsafter implantation. At the end of the study period 41 patients concluded the follow-up.

Results:

After statisticalanalysis, the results demonstrated a favorable electrical performance, with stable chronic values.

Conclusion:

Thelead above is safe and effective for use in routine pacemaking practice.
Subject(s)
Full text: 1 Index: LILACS Main subject: Pacemaker, Artificial / Electrodes, Implanted / Atrioventricular Block / Cost-Effectiveness Analysis Type of study: Prognostic_studies Limits: Adult / Aged / Aged80 / Female / Humans / Male Language: Pt Journal: RELAMPA, Rev. Lat.-Am. Marcapasso Arritm Journal subject: CARDIOLOGIA Year: 2014 Type: Article
Full text: 1 Index: LILACS Main subject: Pacemaker, Artificial / Electrodes, Implanted / Atrioventricular Block / Cost-Effectiveness Analysis Type of study: Prognostic_studies Limits: Adult / Aged / Aged80 / Female / Humans / Male Language: Pt Journal: RELAMPA, Rev. Lat.-Am. Marcapasso Arritm Journal subject: CARDIOLOGIA Year: 2014 Type: Article