Your browser doesn't support javascript.
loading
Análise biomecânica de três tipos de conexões protéticas em implantes dentários de diâmetro reduzido / Biomechanical analysis of three types of prosthetic connections on reduced diameter dental implants
Oliveira, Mayara Moya de; Matsumoto, Mariza Akemi; Butignon, Luís Eduardo; Cardoso, Camila Lopes; Nary Filho, Hugo.
  • Oliveira, Mayara Moya de; Universidade do Sagrado Coração/USC. BR
  • Matsumoto, Mariza Akemi; Universidade do Sagrado Coração/USC. BR
  • Butignon, Luís Eduardo; Universidade do Sagrado Coração/USC. BR
  • Cardoso, Camila Lopes; Universidade do Sagrado Coração/USC. BR
  • Nary Filho, Hugo; Universidade do Sagrado Coração/USC. BR
ImplantNews ; 11(5): 595-599, 2014. tab
Article in Portuguese | LILACS, BBO | ID: lil-731522
RESUMO

Objetivos:

avaliar o comportamento biomecânico de três tipos de conexão protética em implantes de diâmetro reduzido. Material e

Métodos:

implantes com diâmetro 3,25 mm (plataforma 3,4 mm) foram inseridos em polioximetileno (poliacetal) com torque máximo de 45 Ncm e receberam componentes protéticos calcináveis divididos em três grupos G1 cone-morse (CM), hexágono interno (HI), e Hexágono externo (HE). Todo os pilares foram sobrefundidos em liga de CoCr e aparafusados com torque de 20 Ncm conforme recomendações do fabricante. A avaliação biomecânica ocorreu conforme a norma ISO 14801-2007 carregamento compressivo 30º e fadiga cíclica (f=15Hz, 5 milhões de ciclos) em ambiente seco. A análise estatística foi conduzida usando-se os testes Anova e Tukey (nível de significância 5%).

Resultados:

os valores compressivos médios foram G1 (317,43 N), G2 (559,44 N), e G3 (290,83 N). Diferenças signifi cativas foram vistas entre G1 e G2 (p < 0,001) e entre G2 e G3 (p < 0,0001). Não houve diferença entre grupos no teste de fadiga. Os valores correspondentes foram G1 (190,80 N), G2 (163,25 N), e G3 (141,77 N).

Conclusões:

todos os grupos demonstraram desempenho biomecânico aceitável
ABSTRACT

Objectives:

to evaluate the biomechanical behavior of three types of prosthetic abutment connections in reduced implant diameters. Material and

Methods:

3.25 mm-diameter dental implants (3.4 mm implant platform) were inserted in polyoxymethylene (polyacetal) resin with maximum torque of 45Ncm and received burnout prosthetic abutments divided into three groups G1 cone-morse (CM), Internal Hex (IH), and External Hex (EH). All abutments were overcast in a CoCr alloy and fastened to 20 Ncm following manufacturer’s instructions. Biomechanical evaluation was performed according to ISO 14801-2007 standards 30º compressive and fatigue loading (f = 15 Hz, 5 million cycles) in dry environment. Statistical analysis was conducted with Anova and post-hoc Tukey tests (5% signifi cance level).

Results:

mean compressive values were G1 (317.43 N), G2 (559.44 N), and G3 (290.83 N). Signifi cant differences were seen between G1 and G2 (p < 0.001) and between G2 and G3 (p < 0.0001). No differences among groups were seen for fatigue loading. The correspondent endurance limits were G1 (190.80 N), G2 (163.25 N), and G3 (141.77 N).

Conclusions:

all groups demonstrated acceptable biomechanical performance.
Subject(s)

Search on Google
Index: LILACS (Americas) Main subject: Dental Implants / Dental Implant-Abutment Design Language: Portuguese Journal: ImplantNews Journal subject: Dentistry Year: 2014 Type: Article Affiliation country: Brazil Institution/Affiliation country: USC+BR

Similar

MEDLINE

...
LILACS

LIS

Search on Google
Index: LILACS (Americas) Main subject: Dental Implants / Dental Implant-Abutment Design Language: Portuguese Journal: ImplantNews Journal subject: Dentistry Year: 2014 Type: Article Affiliation country: Brazil Institution/Affiliation country: USC+BR