Estudio de eficiencia y sensibilidad de alarmas de dos analizadores hematológicos en un hospital pediátrico / Study of efficiency and sensitivity of alarms of two hematology analyzers in a pediatric hospital / Estudo de eficiência e sensibilidade o alarmes de dois analisadores de hematologia em um hospital pediátrico
Osta, Viviana; Segura, Celia; Tissera, Gladys; Ayuso, Claudia.
  • Osta, Viviana; Hospital de Niños Dr. Ricardo Gutiérrez. Laboratorio Central. Sección Hemocitología. Ciudad Autónoma de Buenos Aires. AR
  • Segura, Celia; Hospital de Niños Dr. Ricardo Gutiérrez. Laboratorio Central. Sección Hemocitología. Ciudad Autónoma de Buenos Aires. AR
  • Tissera, Gladys; Hospital de Niños Dr. Ricardo Gutiérrez. Laboratorio Central. Sección Hemocitología. Ciudad Autónoma de Buenos Aires. AR
  • Ayuso, Claudia; Hospital de Niños Dr. Ricardo Gutiérrez. Laboratorio Central. Sección Hemocitología. Ciudad Autónoma de Buenos Aires. AR
Acta bioquím. clín. latinoam ; 48(1): 0-0, mar. 2014. tab
Article in Spanish | LILACS | ID: lil-734220
RESUMEN
Con el objetivo de mejorar el flujo de trabajo en el Laboratorio de Hemoci-tología se evaluó el desempeño analítico y eficiencia del sistema de alarmas de dos contadores hematológicos de última generación, Coulter LH750 y Abbott CEL-DYN Ruby, para la resolución de muestras patológicas en una población de pacientes pediátricos, considerando como método de referencia la revisión microscópica y la realización del recuento diferencial leucocitario en forma manual de los extendidos de sangre periférica. Se procesaron 178 muestras seleccionadas de la carga de trabajo diario del Laboratorio en ambos contadores, se analizaron los reportes emitidos y se realizó el diferencial manual. Se encontró una excelente concordancia entre los dos instrumentos para el recuento de leucocitos, eritrocitos, hemoglobina, VCM y plaquetas, y entre el diferencial manual y el reportado por ambos equipos. La eficiencia de las alarmas en ambos analizadores fue similar, 77% para CELL-DYN Ruby y 71,3% para LH750. La tasa de falsos negativos fue de 3,4% y 4,0% respectivamente. La tasa de falsos positivos en ambos equipos fue 17%. Se puede concluir que los analizadores evaluados presentan un desempeño comparable. Cada laboratorio debe considerar las características de la población de pacientes que recibe para establecer criterios estrictos y sistemáticos para la revisión del frotis.
ABSTRACT
The goal of the hematology laboratory is to improve workflow efficiency by optimizing instrument results and flagging sensitivity with its daily manual blood smears requirements. Two hematology analyzers (Coulter LH750 and CEL-DYN Ruby Abbott) were evaluated to determine analytical performance and efficiency in daily work at the Central Laboratory of a Pediatric Hospital. Microscopic revision was considered the reference method. A total of 178 pathological blood samples randomly selected from ambulatory and in-patients were analyzed in both analyzers. An excellent correlation between both instruments for leukocytes, erythrocytes, hemoglobin, MCV and platelet count was observed; also close agreement of results between manual differential and automated count. The efficiency of alarms triggered by the automated hematology analyzer was similar. Total efficiency of Cell-Dyn Ruby was 77% and 71.3% in the LH750. The false negative rate was 3.4% and 4.0% for the LH750 and Ruby respectively. The false positive rate was 17% for both instruments. It can be concluded that the two analyzers tested have a comparable analytical performance and each clinical laboratory must consider the characteristics of their population in order to establish critical and systematic criteria for manual microscopic review.
RESUMO
A fim de melhorar o fluxo de trabalho no Laboratòrio de Hematologia avaliou-se o desempenho analítico e a eficiencia do sistema de alarmes dos dois contadores de hematologia de última geragáo, Coulter LH750 e Abott CEL-DYN Ruby, para a resolugáo de amostras patológicas numa populagáo de pacientes pediatricos. Considerou-se como método de referencia a revisáo microscòpica e diferencial manual dos esfregagos de sangue periférico. Foram processadas 178 amostras seleccionadas da carga de trabalho diàrio do Laboratòrio em ambos contadores, foram analisados os relatórios emitidos e se realizou o diferencial manual. Encontrou-se uma excelente concordancia entre os dois instrumentos para a contagem de leucócitos, eritrocitos, hemoglobina, VCM e plaquetas, e entre el diferencial manual e o relatado por ambas equipes. A eficiencia dos alarmes em ambos os analisadores foi semelhante, 77% para CELL-DYN Ruby e 71,3% para LH750. A taxa de falsos negativos foi de 3,4% e 4,0%, respectivamente. A taxa de falsos positivos foi de 17% para ambos equipes. Pode-se concluir que os analisadores avaliados tem um desempenho comparàvel. Cada laboratòrio deve considerar as características da populagáo de pacientes que recebe para estabelecer critérios rigorosos e sistemáticos para a revisáo de esfregagos.
Subject(s)

Full text: Available Index: LILACS (Americas) Main subject: Cost Efficiency Analysis / Proof of Concept Study / Hematologic Tests / Hospitals, Pediatric Type of study: Diagnostic study / Practice guideline / Prognostic study Limits: Humans Language: Spanish Journal: Acta bioquím. clín. latinoam Journal subject: Bioqu¡mica / Qu¡mica Cl¡nica Year: 2014 Type: Article Affiliation country: Argentina Institution/Affiliation country: Hospital de Niños Dr. Ricardo Gutiérrez/AR

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Main subject: Cost Efficiency Analysis / Proof of Concept Study / Hematologic Tests / Hospitals, Pediatric Type of study: Diagnostic study / Practice guideline / Prognostic study Limits: Humans Language: Spanish Journal: Acta bioquím. clín. latinoam Journal subject: Bioqu¡mica / Qu¡mica Cl¡nica Year: 2014 Type: Article Affiliation country: Argentina Institution/Affiliation country: Hospital de Niños Dr. Ricardo Gutiérrez/AR