Cuantificación simultánea de andrógenos, estrógenos, corticoides y pregnanos mediante cromatografía de gases acoplada a espectrometría de masas / Simultaneous quantification of androgens, estrogens, corticoids and pregnanes thorugh the gas chromatography-mass spectrometry method
Martínez Brito, Dayamín; Correa Vidal, Margarita Teresa; Oropesa Rodríguez, Roberto; González Pérez, Odalys; Ledea Lozano, Oscar Ernesto.
  • Martínez Brito, Dayamín; Instituto de Medicina Deportiva. Laboratorio Antidoping. La Habana. CU
  • Correa Vidal, Margarita Teresa; Instituto de Medicina Deportiva. Laboratorio Antidoping. La Habana. CU
  • Oropesa Rodríguez, Roberto; Instituto de Medicina Deportiva. Laboratorio Antidoping. La Habana. CU
  • González Pérez, Odalys; Instituto de Medicina Deportiva. Laboratorio Antidoping. La Habana. CU
  • Ledea Lozano, Oscar Ernesto; Instituto de Medicina Deportiva. Laboratorio Antidoping. La Habana. CU
Rev. cuba. farm ; 48(4)oct.-dic. 2014. ilus, tab
Article in Spanish | LILACS, CUMED | ID: lil-748771
RESUMEN

INTRODUCCIÓN:

la evaluación de hormonas en pacientes de endocrinología y bioquímica clínica, entre otros, se realiza actualmente por métodos de inmunoensayo con las desventajas inherentes a este tipo de análisis (ejemplo reacciones cruzadas). Con la aplicación de la espectrometría de masas acoplada a cromatografía de gases es posible aumentar la sensibilidad y especificidad de los análisis.

OBJETIVO:

validar un ensayo que permita la cuantificación de 21 hormonas mediante la técnica de cromatografía de gases-espectrometría de masas con propósitos de pesquizaje.

MÉTODOS:

el ensayo se basó en una extracción en fase sólida con columnas DetectabuseTM y posterior extracción líquido-líquido a pH básico. Los derivados trimetilsilil se analizaron con un cromatógrafo de gases Hewlett-Packard 6890 acoplado a un espectrómetro de masas cuadrupolar serie 5973. La columna capilar fue HP Ultra-1 (17 m x 0,20 mm x 0,11 µm). La adquisición fue realizada en modo SIM (monitoreo selectivo de iones).

RESULTADOS:

los resultados del proceso de validación cumplieron con los criterios de aceptación de una manera adecuada correspondiente a un método con propósitos de pesquizaje. El método resultó ser específico para los 21 compuestos estudiados, lineal (r2= 0,900-0,996) y preciso (CV entre 1,4-9,6 por ciento). El rendimiento de extracción se mantuvo entre 69-117 por ciento. La robustez con variación de disolvente mostró recobrados entre 69-139 por ciento.

CONCLUSIONES:

los resultados demuestran que la técnica analítica desarrollada para la cuantificación de los 21 compuestos esteroidales en orina, es exacta, específica, lineal, precisa y robusta para su aplicación con propósitos de pesquizaje en estudios endocrinológicos clínicos(AU)
ABSTRACT

INTRODUCTION:

the evaluation of hormones on endocrinology and clinical biochemistry patients, among others, is presently performed, using inmunoassays with its intrinsic disadvantages such as cross reactions. The application of gas chromatography coupled to mass spectrometry allows increasing the sensitivity and specificity of this analysis.

OBJECTIVE:

to validate an assay for quantitation of 21 hormones by the analytical technique of gas chromatography - mass spectrometry for screening purposes.

METHODS:

the assay followed a solid-phase extraction using DetectabuseTM columns and subsequently liquid-liquid extraction at basic pH. Trimethylsilyl derivatives were evaluated on Hewlett-Packard 6890 gas chromatograph coupled to serial 5973 quadrupolar mass spectrometer. The capillary column was HP Ultra-1 (17 m x 0.20 mm x 0.11 µm). Data was gathered following SIM mode (selective ion monitoring).

RESULTS:

the results of the validation process adequately fulfilled the acceptance criteria for screening purposes. The assay proved to be specific for all studied compounds, linear (r2= 0.900-0.996) and precise (VC 1.4-9.6 percent). The extraction yield was kept at 69-117 percent. The robustness assay with varying dissolvent showed extraction yields ranging 69-139 percent.

CONCLUSIONS:

results of the validation process show that the developed assay to simultaneously quantify 21 steroidal compounds in urine is exact, linear, precise, robust and specific for its application in clinical endocrine studies for screening purposes(AU)
Subject(s)

Full text: Available Index: LILACS (Americas) Main subject: Validation Studies as Topic / Hormone Antagonists Limits: Female / Humans / Male Language: Spanish Journal: Rev. cuba. farm Journal subject: Pharmacy Year: 2014 Type: Article Affiliation country: Cuba Institution/Affiliation country: Instituto de Medicina Deportiva/CU

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Main subject: Validation Studies as Topic / Hormone Antagonists Limits: Female / Humans / Male Language: Spanish Journal: Rev. cuba. farm Journal subject: Pharmacy Year: 2014 Type: Article Affiliation country: Cuba Institution/Affiliation country: Instituto de Medicina Deportiva/CU