Your browser doesn't support javascript.
loading
Aplicación de un modelo de un acelerador Elekta Precise basado en Monte Carlo para la evaluación de la respuesta de detectores de estado sólido / Application of an Elekta Precise lineal accelerator model based on Monte Carlo to evaluate the solid state detectors response
Linares Rosales, Haydee M; Lara Mas, Elier; Alfonso Laguardia, Rodolfo.
  • Linares Rosales, Haydee M; Instituto Superior de Ciencias y Tecnologías Aplicadas, (INTEC). La Habana. CU
  • Lara Mas, Elier; Instituto Superior de Ciencias y Tecnologías Aplicadas, (INTEC). La Habana. CU
  • Alfonso Laguardia, Rodolfo; Instituto Superior de Ciencias y Tecnologías Aplicadas, (INTEC). La Habana. CU
Nucleus (La Habana) ; (57): 13-18, Jan.-June 2015.
Article in English | LILACS | ID: lil-754873
RESUMEN
En el trabajo se realizó el análisis de la respuesta de diferentes detectores de estado sólido a partir de un modelo de Monte Carlo de un acelerador lineal Elekta Precise, para los haces de energías de 6 MV y 15 MV. Para ello se realizaron simulaciones con el código EGSnrc. Se calculó la dosis depositada en un maniquí de agua voxelizado con su superficie a 100 cm de la fuente, empleando los valoresóptimos de energía media y FWHM del haz primario de electrones para este modelo. A partir de la dosis depositada en el maniquí se construyeron las curvas de dosis en profundidad y perfiles de dosis a diferentes profundidades. Las curvas se compararon con valores medidos para cada detector empleado en un arreglo experimental similar a la simulación realizada, aplicando criterios de aceptabilidad basados en intervalos de confianza. De forma adicional se analizó para cada caso la dosis en función del tamaño de campo. Se obtuvo una buena correspondencia entre las simulaciones y las mediciones, encontrándose todos los resultados dentro de los márgenes de tolerancia.
ABSTRACT
The evaluation of the solid state detectors response based on a Monte Carlo model of an Elekta Precise lineal accelerator, was done in this work for the beam energies of 6 y 15 MV. Simulations were performed using the EGSnrc code. Employing the optimal values of mean energy and FWHM from the primary electron beam, deposited dose in a voxelized water phantom at 100 cm of source to surface distance was calculated. Depth dose curves and lateral dose profi les were obtained. Comparison between simulations and the experimental values obtained for each detector, were done using acceptability criteria based on confi dence limits. Additionally outputs factors were analyzed in each one of the study cases. Good agreement between simulations and measurements were reached.

Full text: Available Index: LILACS (Americas) Type of study: Prognostic study Language: English Journal: Nucleus (La Habana) Journal subject: Nuclear Medicine / Radiotherapy Year: 2015 Type: Article Affiliation country: Cuba Institution/Affiliation country: Instituto Superior de Ciencias y Tecnologías Aplicadas, (INTEC)/CU

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Type of study: Prognostic study Language: English Journal: Nucleus (La Habana) Journal subject: Nuclear Medicine / Radiotherapy Year: 2015 Type: Article Affiliation country: Cuba Institution/Affiliation country: Instituto Superior de Ciencias y Tecnologías Aplicadas, (INTEC)/CU