Your browser doesn't support javascript.
loading
Ruptura de electrodo ventricular por síndrome de "twiddler" / Ventricular electrode rupture due to twiddler's syndrome
Díaz, Juan C; Duque, Mauricio; Marín, Jorge E; Velásquez, Jorge E; Uribe, William.
  • Díaz, Juan C; CES Cardiología. Clínica Las Américas. Medellín. CO
  • Duque, Mauricio; CES Cardiología. Clínica Las Américas. Medellín. CO
  • Marín, Jorge E; CES Cardiología. Clínica Las Américas. Medellín. CO
  • Velásquez, Jorge E; CES Cardiología. Clínica Las Américas. Medellín. CO
  • Uribe, William; CES Cardiología. Clínica Las Américas. Medellín. CO
Rev. colomb. cardiol ; 23(3): 228.e1-228.e4, mayo-jun. 2016. ilus
Article in Spanish | LILACS, COLNAL | ID: lil-791281
RESUMEN
El síndrome de "twiddler" se caracteriza por el enrollamiento y desplazamiento de los electrodos de un dispositivo de estimulación cardíaca debido a rotación a lo largo de su eje mayor, ya sea de manera espontánea o por manipulación del dispositivo. En la mayoría de los casos, dicho desplazamiento se acompaña de daño del electrodo; no obstante, no se ha descrito la ruptura total del electrodo. A continuación se describe el caso de un paciente con marcapasos unicameral implantado por fibrilación auricular lenta, en quien se documentó disfunción del mismo con falla en la detección y la captura; al interrogatorio el paciente manifestó astenia, adinamia, intolerancia al ejercicio y presíncope. La evaluación radiológica documentó enrollamiento del electrodo formando asas apretadas, con ruptura completa del electrodo y desplazamiento del generador del marcapasos en un bolsillo amplio. Según criterio de los autores, este es el primer caso de síndrome de twiddler en el que se documenta ruptura completa del electrodo.
ABSTRACT
Twiddler's syndrome is characterised by a twist and displacement of the electrodes of a cardiac stimulation device due to a rotation on its main axis, either spontaneously or when manipulating the pacemaker. In most cases this movement is accompanied by damage of the electrode; however, complete rupture of the lead has not been described. A case of a patient carrying a unicameral pacemaker required due to slow atrial fibrillation is now described, where a dysfunction of the device was documented as a failure to detect and capture. Upon interrogation the patient manifested asthenia, adynamia, exercise intolerance and pre-syncope. Radiological assessment documented a twisted electrode with the shape of tight handles, with complete electrode rupture and displacement of the pacemaker generator in a wide pocket. According to the authors’ criteria, this is the first Twiddler's syndrome case where total rupture of an electrode has been documented.
Subject(s)


Full text: Available Index: LILACS (Americas) Main subject: Biological Clocks Limits: Aged / Aged80 / Humans / Male Language: Spanish Journal: Rev. colomb. cardiol Journal subject: Cardiology Year: 2016 Type: Article Affiliation country: Colombia Institution/Affiliation country: CES Cardiología/CO