The present investigation targets to develop a simple, specific, sensitive and accurate reverse phase high performance liquid chromatographic (RP-HPLC) method in human plasma for the estimation of metformin HCl and propranolol HCl from bulk drug and also from the marketed products. Human plasma samples were subjected to correct procedure for protein precipitation by methanol and protein free plasma samples were directly injected into HPLC C18 column. Chromatographic determination was performed on a reversed phase C18 column (3.9 mm X 300 mm, particle size 5 μm) using a mixture of acetonitrile and 0.1M pH 4.5 potassium dihydrogenortho phosphate buffer (mL) (4060) at a flow rate of 0.8 mL/min and maintained at a pressure of 140 to 150 Kg/cm2. The retention time for metformin HCl and propranolol HCl was found to be 9.084 min and 6.132 min respectively at 232 nm without any interference of endogenous compounds in the plasma. The method was linear in the range between 50-2000 ng/mL. The peak areas were reproducible as indicated by low coefficient of variation. It was found that the excipients in the tablet dosage form do not interfere in the quantification of active drug by proposed method.