Your browser doesn't support javascript.
loading
Patient Dose Assessment for Common Digital Diagnostic Radiology Examination in Hamad Medical Corporation Hospitals in the State of Qatar.
Br J Med Med Res ; 2016; 14(8): 1-7
Article in English | IMSEAR | ID: sea-182855
ABSTRACT
The aim of this study was to estimate radiation doses to patients undergoing standard radiographic examinations using Computed Radiography (CR) and Direct Digital Radiography (DDR) in two hospitals within Hamad Medical Corporation (HMC) in Qatar, and compare the results with regional and international Diagnostic Reference Levels (DRLs). Data on 3391 patients were recorded from different X-ray rooms in HMC hospitals. Entrance Skin Dose (ESD) was measured for 1046 patients for the most five common X-ray examinations (a total of 7 projections) namely Skull, Chest, Abdomen, Lumbar Spine and Pelvis. Exposure factors such as kV, mAs and Focal to Skin Distance (FSD) were recorded for each patient. Tube Output was measured for a range of selected kV values. ESD for each individual patient was calculated using the tube output and the technical exposure factors for each examination. The ESD values were compared with some international DRLs for all types of examinations. HMC patient demographic data evaluated from this study were average age of 39 years, average weight of 60-80 kg and mean height of 165 cm. The most procedure performed during the time of this study was chest PA (52%), and the least procedure performed was skull AP/LAT (1%) examination. The mean ESD values found to be generally lower than the published values. With exception of abdomen examinations at Hamad General Hospital (HGH), mean ESD values were found to be within the established IAEA (DRL). The mean ESD values at HGH were found to be much higher than that at Al Khor Hospital (AKH) for abdomen, Lumbar spine AP, Lumbar Spine LAT and pelvis, but generally lower than the published values.

Full text: Available Index: IMSEAR (South-East Asia) Type of study: Diagnostic study Language: English Journal: Br J Med Med Res Year: 2016 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: IMSEAR (South-East Asia) Type of study: Diagnostic study Language: English Journal: Br J Med Med Res Year: 2016 Type: Article