Comparison between computerized tomography-guided bronchial width measurement versus conventional method for selection of adequate double lumen tube size
Ann Card Anaesth ; 2019 Oct; 22(4): 358-364
Article | IMSEAR | ID: sea-185867
ABSTRACT

Background:

Selection of adequate size double lumen tube (DLT) is complicated by marked inter-individual variability in morphology and dimensions of tracheobronchial tree. Computerized tomography (CT)-guided left bronchus width measurement has been used to predict adequate size DLT in European and Singapore population; however, no such data exist for Indian population who are racially different. We compared the effect of DLT size selection based on CT-guided bronchial width measurement to the conventional method of DLT selection on the adequacy of both lungs isolation and on the safety margin of right-sided DLT.

Methods:

Fifty-five adults scheduled to undergo thoracotomy were enrolled in this prospective observational study. An appropriate size left- or right-sided DLT with outer diameter 0.5–1 mm smaller than the CT-measured bronchial width was selected for the isolation of lungs. Adequacy of separation was checked using fiberoptic bronchoscope. The safety margin of selected right-sided DLT size was calculated from CT-measured right upper lobe bronchus width and diameter of right upper lobe ventilation slot of the DLT.

Results:

Adequate separation of lungs was achieved in 92.7% of studied population, 90.9% in males, and 95.4% in females. Among these, 54.9% patients required different sized DLT as compared to conventional method. Overall safety of margin of right-sided DLTs was comparable between two methods of DLT selection (median [IQR] 4.8 (3.5–6.8) vs. 6.59 (3.5–7.8), P = 0.317). DLT size with adequate isolation of lung correlated with height, tracheal width (TW) on chest X-ray, and age of the patients. A formula to calculate DLT size based on these variable was derived.

Conclusion:

CT-measured bronchial width predicts the appropriate DLT size better than conventional method. In the absence of CT scan facility, patient height, age, and chest X-ray TW may be used to predict DLT size with reasonable accuracy.