LC-MS method development for the quantitation of potential genotoxic impurity 2-Methyl-6-nitro aniline in Telmisartan API
Article | IMSEAR | ID: sea-210753
ABSTRACT
Current work discloses the sensitive LC-MS/MS method development for the trace level determination of genotoxicimpurity 2-Methyl-6-nitro aniline in Telmisartan. 2-Methyl-6-nitro aniline was determined by LC-MS/MS methodin selected ion monitoring mode using LiChrospher RP-18 (100 × 4.6 mm) 5.0 µm column. Gradient technique wasapplied for the elution of analytes using acetonitrile (mobile phase A) and 0.01 M ammonium acetate buffer (mobilephase B) in different ratios. The gradient program (T/%B) was set as 0/5, 2.50/15, 5.00/30, 10.00/50, 15.00/95, and20.00/95. Developed method was validated as per International Conference on Harmonization guidelines. The limit ofdetection and limit of quantitation values found for 2-Methyl-6-nitro aniline were 0.05 and 0.1 µg/ml. The developedmethod serves as an upright tool in quality control for quantitation of 2-Methyl-6-nitro aniline impurity at trace levelsin Telmisartan.

Full text: Available Index: IMSEAR (South-East Asia) Year: 2020 Type: Article