Your browser doesn't support javascript.
loading
External quality assessment survey for hematological laboratories in China.
Southeast Asian J Trop Med Public Health ; 1999 ; 30 Suppl 3(): 1-5
Article in English | IMSEAR | ID: sea-35855
ABSTRACT
To evaluate the feasibility of improving external quality assessment (EQA), we set up an experimental EQA survey that included 465 participants in China. During the period of this survey, we checked the quality of the EQA samples, divided the participants into different groups, each laboratory's result was assessed by calculation standard deviation index (SDI). The reference values were determined to evaluate the accuracy for peer groups. The data showed that the stability of the EQA samples was acceptable. Except for WBC count of the Abbott group, the mean, median and reference values for each parameter were very close. We found that the main reason affecting the performance of the participants was not using the reagents. calibrator and QC material recommended by manufacturer. From this survey, we obtain a good reference for future improvement.
Subject(s)
Full text: Available Index: IMSEAR (South-East Asia) Main subject: Quality Assurance, Health Care / Reference Standards / Reference Values / Humans / Calibration / China / Peer Review, Health Care / Hematologic Tests / Laboratories Country/Region as subject: Asia Language: English Journal: Southeast Asian J Trop Med Public Health Year: 1999 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: IMSEAR (South-East Asia) Main subject: Quality Assurance, Health Care / Reference Standards / Reference Values / Humans / Calibration / China / Peer Review, Health Care / Hematologic Tests / Laboratories Country/Region as subject: Asia Language: English Journal: Southeast Asian J Trop Med Public Health Year: 1999 Type: Article