Your browser doesn't support javascript.
loading
Measurement error of spiral CT Volumetry: Influence of Low Dose CT Technique
Journal of the Korean Radiological Society ; : 69-75, 2004.
Article in English | WPRIM | ID: wpr-101158
ABSTRACT

PURPOSE:

To examine the possible measurement errors of lung nodule volumetry at the various scan parameters by using a small nodule phantom. MATERIALS AND

METHODS:

We obtained images of a nodule phantom using a spiral CT scanner. The nodule phantom was made of paraffin and urethane and its real volume was known. For the CT scanning experiments, we used three different values for both the pitch of the table feed, i.e. 11, 115 and 12, and the tube current, i.e. 40 mA, 80 mA and 120 mA. All of the images acquired through CT scanning were reconstructed three dimensionally and measured with volumetry software. We tested the correlation between the true volume and the measured volume for each set of parameters using linear regression analysis.

RESULTS:

For the pitches of table feed of 11, 11.5 and 12, the mean relative errors were 23.3%, 22.8% and 22.6%, respectively. There were perfect correlations among the three sets of measurements (Pearson's coefficient = 1.000, p<0.001). For the tube currents of 40 mA, 80 mA and 120 mA, the mean relative errors were 22.6%, 22.6% and 22.9%, respectively. There were perfect correlations among them (Pearson's coefficient = 1.000, p<0.001).

CONCLUSION:

In the measurement of the volume of the lung nodule using spiral CT, the measurement error was not increased in spite of the tube current being decreased or the pitch of table feed being increased.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Paraffin / Urethane / Tomography, X-Ray Computed / Linear Models / Tomography, Spiral Computed / Lung Language: English Journal: Journal of the Korean Radiological Society Year: 2004 Type: Article