Your browser doesn't support javascript.
loading
Axial Length and Anterior Chamber Depth by IOL-Master, A-scan according to Viscosity of Silicone Oil
Journal of the Korean Ophthalmological Society ; : 261-266, 2008.
Article in Korean | WPRIM | ID: wpr-112567
ABSTRACT

PURPOSE:

To evaluate differences between partial coherence laser interferometry (IOL-Master, Zeiss) and A-scan measurement of axial length and anterior chamber depth in silicone oil-filled eyes according to viscosity.

METHODS:

Using IOL-Master and A-scan, axial length and anterior chamber depth in silicone oil-filled eyes (n=54) and normal eyes (control, n=54) were measured and analyzed. In silicone oil-filled eyes, calculated axial lengths by A-scan using conversion factors, axial length multiplied by 0.71, and vitreous cavity multiplied by 0.64 (classic method) were compared with those calculated by IOL-Master. Anterior chamber depths were also analyzed., and axial lengths and anterior chamber depths were compared according to the viscosities of silicone oil for measurement by A-scan.

RESULTS:

Axial length and anterior chamber depth using IOL-Master were shorter than those using A-scan by 9.45+/-1.81 mm (p<0.05) and 0.11+/-1.29 mm, respectively. In normal eyes, axial length and anterior chamber depth using IOL-Master and A-scan were not significantly different. In silicone oil-filled eyes, axial length using IOL-Master and conversion factor was also not significantly different. At the highest silicone oil viscosity the difference in measured axial length was greatest (p<0.05) while the difference in anterior chamber depths was smallest.

CONCLUSIONS:

In silicone oil-filled eyes, axial length by IOL-Master was more accurate than that by A-scan, regardless of silicone oil viscosity. Thus, IOL-Master is more useful than A-scan when measuring axial length in silicone oil-filled eyes.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Viscosity / Silicone Oils / Eye / Interferometry / Anterior Chamber Language: Korean Journal: Journal of the Korean Ophthalmological Society Year: 2008 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Viscosity / Silicone Oils / Eye / Interferometry / Anterior Chamber Language: Korean Journal: Journal of the Korean Ophthalmological Society Year: 2008 Type: Article