Selective pulsed dye laser thrombus ablation : Can contrast media replace fiberoptics in pulsed dye laser delivery system?
Korean Circulation Journal ; : 408-414, 1993.
Article in Korean | WPRIM | ID: wpr-115430
ABSTRACT

BACKGROUND:

The absorption coefficient of the arterial thrombus is 7~10 fold greater than that of arterial wall for the wave lengths of 480~510nm. This fact suggest that selective laser thrombus ablation could be possible without ablation of surrounding vessel wall in this wavelengths.

METHODS:

Pulsed dye laser(488nm, 2.2micros, 10Hz) thrombus ablation was investigated in viro. Ablation threshods of whole blood clot and normal aortic tissue were measured using a 1.5mm multifiber laser catheter(14x150microm). Whole blood clot samples were lasered to determine the ablation efficiency. The operative fluence was 80mJ/mm2. The transmission efficiencies of 488nm wave length through various transmission media(including contrast media of saline solution) were observed using a upsilon Fr arterial sheath filled with transmission medium.

RESULTS:

The ablation threshold for fresh whole blood clot in saline solution was 36.5+/-2.4mJ/mm2(9.0+/-0.6mJ/pulse) as compared with that for normal human aortic tissue which was over 405mJ/mm2(>100mJ/pulse)(p<0.05). Pulsed dye laser ablated whole blood clot with the efficiency at 0.026+/-0.009gm/J. The role of thrombus ablation was 0.154+/-0.056gm/min. This experiment also revealed that cotrast media(Ultravist(R) 370, Schering, France) allows light tramission of 488nm wave length at relatively high efficiency.

CONCLUSIONS:

These facts suggested that selective pulsed dye laser ablation of fresh intravascular thrombi can be obtained at relatively low energy fluence without damaging the surrounding vascular tissue. A new pulse dye laser delivery system such as catheter system or native coronary artery filled with contrast media, which has large area of frontfiring face, may facilitate a safe and effective thrombus ablation.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Thrombosis / Sodium Chloride / Contrast Media / Coronary Vessels / Absorption / Lasers, Dye / Catheters Limits: Humans Language: Korean Journal: Korean Circulation Journal Year: 1993 Type: Article