Your browser doesn't support javascript.
loading
Basic Evaluation of Roche OMNI 9 Blood Gas Analyzer / 대한진단검사의학회지
The Korean Journal of Laboratory Medicine ; : 153-158, 2002.
Article in Korean | WPRIM | ID: wpr-121988
ABSTRACT

BACKGROUND:

Recent blood gas analyzers offer electrolytes, Hb-derivatives (O2 Hb, HHb, COHb, MetHb and SulfHb), content of total hemoglobin (ctHb), and metabolites (glucose and lactate) as well as conventional blood gas parameters. Evaluation and efficient control of these instruments affect greatly monitoring and therapy of patients. The authors evaluated a recently introduced blood gas analyzer, the Roche OMNI 9 (Roche Diagnostics) for precision, analytic speed, function, features and comparability with other instruments according to NCCLS guidelines to see if it was suitable for the stat analyzer.

METHODS:

Commercial control materials, AUTO-trol PLUS Level 1, 2, 3 (Roche Diagnostics) were analyzed to evaluate the within-day precision, the between-day precision, and patient samples were used to evaluate comparability with other instruments.

RESULTS:

The within-day and the between-day precisions of the Roche OMNI 9 showed a very low CV of 0- 3.93% and 0- 3.99%, respectively, in all parameters except the lactate, which had moderate CV values of 7.74% and 5.93% in Level 3 QC material (low concentration). In compari-son with the Bayer instrument, correlation was high in all parameters with the r(2) value ranging from 0.9441 to 0.9901 except for COHb (r(2)=0.4239) and MetHb (r(2)=0.0034) whose concentrations were in the lower level. Glucose (r(2)=0.991), lactate (r(2)=0.9824), and hemoglobin (r(2)=0.9961) showed a rather high correlation. The Roche OMNI 9 could analyze 40 samples of blood gas and 27 samples of full parameter per hour, which was comparable to the Bayer M855.

CONCLUSIONS:

The Roche OMNI 9 showed high precision and rapid turnaround time for blood gas and in a selection of glucose, lactate and hemoglobin.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Blood Gas Analysis / Lactic Acid / Electrolytes / Glucose Type of study: Practice guideline Limits: Humans Language: Korean Journal: The Korean Journal of Laboratory Medicine Year: 2002 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Blood Gas Analysis / Lactic Acid / Electrolytes / Glucose Type of study: Practice guideline Limits: Humans Language: Korean Journal: The Korean Journal of Laboratory Medicine Year: 2002 Type: Article