Your browser doesn't support javascript.
loading
Comparison of Keratometry and Corneal Higher Order Aberrations between Scout Videokeratoscope and Pentacam Scheimpflug Camera
Journal of the Korean Ophthalmological Society ; : 1758-1764, 2014.
Article in Korean | WPRIM | ID: wpr-140820
ABSTRACT

PURPOSE:

To investigate the correlations of keratometry and corneal high order aberrations (HOAs) between the Keratron Scout videokeratoscope and the Pentacam HR Scheimpflug camera.

METHODS:

From December 2012 to February 2013, keratometry and corneal HOAs were determined using the Keratron Scout videokeratoscope and Pentacam HR Scheimpflug camera in 23 healthy individuals (46 eyes).

RESULTS:

Average keratometry showed high correlation with 95% confidence interval -0.155 +/- 0.37 between the Keratron Scout and Pentacam HR. When comparing HOAs of the Keratron Scout and total corneal HOAs of Pentacam HR, total root-mean-square (RMS), and spherical aberration were positively correlated between the 2 devices (r = 0.67, p < 0.001 and r = 0.74, p < 0.001, respectively). When comparing HOAs of Keratron Scout and anterior corneal HOAs of Pentacam HR, total RMS, spherical aberration and coma were positively correlated between the 2 devices (r = 0.62, p < 0.001, r = 0.81, p < 0.001, and r = 0.52, p = 0.047, respectively).

CONCLUSIONS:

Although the 2 devices are based upon different principles, the Keratron Scout videokeratoscope and Pentacam HR Scheimpflug camera showed positive correlations in keratometry, total RMS, spherical aberration, and coma. Both devices may be useful for clinical applications.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Coma Language: Korean Journal: Journal of the Korean Ophthalmological Society Year: 2014 Type: Article