Your browser doesn't support javascript.
loading
Platelet Distribution Width from Two Automated Hematology Analysers: A Correlation Analysis
Journal of Laboratory Medicine and Quality Assurance ; : 42-46, 2017.
Article in Korean | WPRIM | ID: wpr-156755
ABSTRACT
Platelet distribution width (PDW) is an index for platelet size variation. In this study, we analysed the correlation between PDW values obtained using two different hematology analysers that employ different measurement methods. Complete blood cell parameters including PDW for 153 healthy individuals were measured using both, ADVIA 2120i (Simens AG, Germany) and XN-3000 (Sysmex, Japan). The PDW values measured using the two hematology analysers showed a moderate correlation (r=0.661, P0.900, P<0.001). PDW obtained using XN-3000 showed a strong correlation with mean platelet volume, whereas PDW obtained using ADVIA 2120i did not. The reference values in this group were 40.0%–64.2% in ADVIA 2120i and 9.0–16.0 fL in XN-3000. In conclusion, PDW values obtained using ADVIA 2120i and XN-3000 are not interchangeable. In laboratories equipped with more than one hematology analyser, a particular analyser should be used consistently for monitoring a particular patient.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Platelet Count / Reference Values / Blood Cells / Blood Platelets / Erythrocyte Indices / Erythrocytes / Mean Platelet Volume / Hematology / Leukocytes Limits: Humans Language: Korean Journal: Journal of Laboratory Medicine and Quality Assurance Year: 2017 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Platelet Count / Reference Values / Blood Cells / Blood Platelets / Erythrocyte Indices / Erythrocytes / Mean Platelet Volume / Hematology / Leukocytes Limits: Humans Language: Korean Journal: Journal of Laboratory Medicine and Quality Assurance Year: 2017 Type: Article