Development and validation of a UPLC-MS/MS method for the quantification of acetaminophen in human plasma and its application to pharmacokinetic studies
Translational and Clinical Pharmacology
;
: 30-36, 2016.
Article
in English
| WPRIM
| ID: wpr-165360
ABSTRACT
We developed an ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) method for the determination of acetaminophen concentration in human plasma. Following protein precipitated extraction, the analytes were separated and analyzed using an UPLC-MS/MS in the multiple reaction monitoring (MRM) mode with the respective [M+H]+ ions, m/z 152.06 → 110.16 for acetaminophen and m/z 180.18 → 138.12 for phenacetin (internal standard, IS). The method showed a linear response from 1 to 100 µg/mL (r > 0.9982). The limit of quantitation for acetaminophen in plasma was 1 µg/mL. The intra- and inter-day accuracy ranged in the ranges of 94.40–99.56% and 90.00–99.20%, respectively. The intra- and inter-day precision ranged in the ranges of 2.64–10.76% and 6.84–15.83%, respectively. This method was simple, reliable, precise and accurate and can be used to determine the concentration of acetaminophen in human plasma. Finally, this fully validated method was successfully applied to a pharmacokinetic study of acetaminophen in healthy volunteers following oral administration.
Full text:
Available
Index:
WPRIM (Western Pacific)
Main subject:
Plasma
/
Mass Spectrometry
/
Phenacetin
/
Administration, Oral
/
Healthy Volunteers
/
Ions
/
Acetaminophen
Limits:
Humans
Language:
English
Journal:
Translational and Clinical Pharmacology
Year:
2016
Type:
Article
Similar
MEDLINE
...
LILACS
LIS