Finite element study on the effect of abutment length and material on implant bone interface against dynamic loading
The Journal of Advanced Prosthodontics ; : 140-144, 2011.
Article in English | WPRIM | ID: wpr-172523
ABSTRACT

PURPOSE:

Finite element study on the effect of abutment length and material on implant bone interface against dynamic loading. MATERIALS AND

METHODS:

Two dimensional finite element models of cylinderical implant, abutments and bone made by titanium or polyoxymethylene were simulated with the aid of Marc/Mentat software. Each model represented bone, implant and titanium or polyoxymethylene abutment. Model 1 Implant with 3 mm titanium abutment, Model 2 Implant with 2 mm polyoxymethylene resilient material abutment, Model 3 Implant with 3 mm polyoxymethylene resilient material abutment and Model 4 Implant with 4 mm polyoxymethylene resilient material abutment. A vertical load of 11 N was applied with a frequency of 2 cycles/sec. The stress distribution pattern and displacement at the junction of cortical bone and implant was recorded.

RESULTS:

When Model 2, 3 and 4 are compared with Model 1, they showed narrowing of stress distribution pattern in the cortical bone as the height of the polyoxymethylene resilient material abutment increases. Model 2, 3 and 4 showed slightly less but similar displacement when compared to Model 1.

CONCLUSION:

Within the limitation of this study, we conclude that introduction of different height resilient material abutment with different heights i.e. 2 mm, 3 mm and 4 mm polyoxymethylene, does not bring about significant change in stress distribution pattern and displacement as compared to 3 mm Ti abutment. Clinically, with the application of resilient material abutment there is no significant change in stress distribution around implant-bone interface.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Titanium / Dental Implants / Finite Element Analysis / Displacement, Psychological Language: English Journal: The Journal of Advanced Prosthodontics Year: 2011 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Titanium / Dental Implants / Finite Element Analysis / Displacement, Psychological Language: English Journal: The Journal of Advanced Prosthodontics Year: 2011 Type: Article