Your browser doesn't support javascript.
loading
Construction and Performance Evaluation of Digital Radiographic System / 의학물리
Korean Journal of Medical Physics ; : 144-148, 2007.
Article in Korean | WPRIM | ID: wpr-177493
ABSTRACT
Current digital radiography systems are rapidly growing in clinical applications. The purpose of this study was to evaluate the characteristics of a mobile digital radiographic system. The performance of the mobile DR system was evaluated by measuring the modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE). Measurements were made on a LISTEM Mobix-1000 generator and a Teleoptic PRA Alpha-R4000 detector. Imaging characteristics were measured for these two systems using the IEC-61267 defined RQA5 (kVp 74, additional filtration 21 mmAl) radiographic condition. The MTF at 10% was measured as 2.4 cycles/mm and the DQE(0) values for radiation exposure 0.19, 0.5, and 1.3 mR were measured as 54%, 55%, and 76%, respectively. The NPS curves gradually decreased at high spatial frequencies. This high DQE at low frequencies, may be useful for low frequency information. The results suggested that mobile DR system could be integrated with emergency ambulance system in teleradiologic imaging applications.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Radiographic Image Enhancement / Ambulances / Emergencies / Filtration / Noise Language: Korean Journal: Korean Journal of Medical Physics Year: 2007 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Radiographic Image Enhancement / Ambulances / Emergencies / Filtration / Noise Language: Korean Journal: Korean Journal of Medical Physics Year: 2007 Type: Article