Your browser doesn't support javascript.
loading
Neutron Generation from a 24 MV Medical Linac / 의학물리
Korean Journal of Medical Physics ; : 97-103, 2005.
Article in Korean | WPRIM | ID: wpr-187031
ABSTRACT
The energy spectra and dose calculations were performed for secondary neutrons from a 24 MV LINAC using MCNPX code (V2, 4, 0). The energy spectra for neutrons and photons emitted from the LINAC head, and absorbed dose to water were calculated in water phantom. The absorbed doses calculated with Monte Carlo were 0.66~0.35 mGy/photon Gy at the surface to d=5 cm, and calculated with interaction data was 0.52 mGy/photon Gy at the depth of electron equilibrium in water. We have shown that this work can be applied to dose estimation of neutrons from high energy LINAC through the comparison of our results with other results.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Water / Photons / Head / Neutrons Language: Korean Journal: Korean Journal of Medical Physics Year: 2005 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Water / Photons / Head / Neutrons Language: Korean Journal: Korean Journal of Medical Physics Year: 2005 Type: Article