Your browser doesn't support javascript.
loading
Intraocular Lens Power Calculations Using Dual Scheimpflug Analyzer
Journal of the Korean Ophthalmological Society ; : 369-379, 2016.
Article in Korean | WPRIM | ID: wpr-189735
ABSTRACT

PURPOSE:

To investigate the accuracy of intraocular lens power calculations using simulated keratometry (simK) of dual Scheimpflug analyzer and 5 types of formulas in cataract patients.

METHODS:

The keratometry (K), axial length (AXL) and anterior chamber depth (ACD) were measured using ultrasound biometry (USB) combined with auto-keratometry (Auto-K), parital coherence interferometry (PCI; IOL master®) and dual Scheimpflug analyzer (DSA; Galilei®) in 39 eyes of 39 patients. Predicted refraction was calculated using Auto-K, mean K of PCI, and simK and total corneal power (TCP) of DSA in the Sanders-Retzlaff-Kraff (SRK-T) formula. The SRK-II, SRK-T, Holladay II, Haigis, and Hoffer-Q formula were used to calculate predicted refraction with the simK of DSA and AXL of USB. Manifest refraction, mean numerical error (MNE) and mean absolute error were evaluated 1, 3 and 6 months after cataract surgery.

RESULTS:

TCP of DSA was lower compared with other keratometric values (p < 0.05). The MNE was not different among Auto-K, mean K and simK. The MNE using TCP was larger compared with Auto-K, mean K and simK at 1 month after surgery (p < 0.05). There was a difference in MNE between simK and TCP of DSA at 6 months after surgery (p < 0.05). The MNE of SRK-T formula was the smallest in the intraocular lens (IOL) power calculation using the simK of DSA.

CONCLUSIONS:

We suggest using IOL power calculations with simK of DSA and SRK-T formula rather than TCP of DSA in cataract patients with normal corneas.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Cataract / Biometry / Ultrasonography / Cornea / Interferometry / Lenses, Intraocular / Anterior Chamber Type of study: Diagnostic study Limits: Humans Language: Korean Journal: Journal of the Korean Ophthalmological Society Year: 2016 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Cataract / Biometry / Ultrasonography / Cornea / Interferometry / Lenses, Intraocular / Anterior Chamber Type of study: Diagnostic study Limits: Humans Language: Korean Journal: Journal of the Korean Ophthalmological Society Year: 2016 Type: Article