ABSTRACT

PURPOSE:

To investigate the relationship between optical coherence tomography (OCT) and scanning laser polarimetry (SLP) in measuring peripapillary retinal nerve fiber layer (RNFL) thickness in glaucomatous eyes.

METHODS:

Fifty glaucomatous eyes were evaluated in this study. Evaluations were analyzed two ways. First, parameters of the Stratus OCT (average thickness, superior/ inferior average) and GDx VCC (TSNIT average, nerve fiber indicator (NFI), superior/ inferior average) were correlated using the Pearson's correlation coefficient (r). Secondly, comparison (r) of these parameters was completed using the mean deviation (MD) of visual field defect.

RESULTS:

The following parameters were found to be significantly correlated (P<0.005). TSNIT average/average thickness (r=0.673), NFI/average thickness (r=-0.742), superior average (r=0.841), and inferior average (r=0.736). In the correlation analysis using the severity of visual field defect, all these parameters had statistically meaningful correlations (P<0.005).

CONCLUSIONS:

GDx VCC and Stratus OCT are highly correlated in glaucomatous eyes. Therefore, peripapillary RNFL thickness measured by Stratus OCT and GDx VCC may be equally helpful in the diagnosis of glaucoma.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Retina / Severity of Illness Index / Glaucoma / Reproducibility of Results / Tomography, Optical Coherence / Laser Scanning Cytometry Limits: Adult / Aged / Aged80 / Female / Humans / Male Language: English Journal: Korean Journal of Ophthalmology Year: 2006 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Retina / Severity of Illness Index / Glaucoma / Reproducibility of Results / Tomography, Optical Coherence / Laser Scanning Cytometry Limits: Adult / Aged / Aged80 / Female / Humans / Male Language: English Journal: Korean Journal of Ophthalmology Year: 2006 Type: Article