ABSTRACT
In vivo testings of the monoleaflet polymer valve were performed in seven dogs to prove its blood biocompatibility. The monoleaflet polymer valve used in this study was developed for short-term usage n the ventricular assist device. The frame and leaflet of the polymer valve were made of polyurethane. The inter-aortic valved conduit were implanted in four dogs and the ventriculo-atrial valved conduit was implanted in one dog. The ventricular assist devices with polymer valve were implanted in two dogs. The longest survival was 20 days. Main causes of death were bleeding and infection. To examine the blood compatibility, each blood sample was collected and RBC, WBC, hematocrit, hemoglobin, platelet and lactic acid dehydrogenase were analyzed. These studies thus far demonstrated that, with further development, a reliable and inexpensive polymer valve will be used in the ventricular assist device as short term usage.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Oxidoreductases / Polymers / Polyurethanes / Blood Platelets / Heart-Assist Devices / Cause of Death / Lactic Acid / Animal Experimentation / Heart Valves / Hematocrit Limits: Animals Language: Korean Journal: The Korean Journal of Thoracic and Cardiovascular Surgery Year: 1997 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Oxidoreductases / Polymers / Polyurethanes / Blood Platelets / Heart-Assist Devices / Cause of Death / Lactic Acid / Animal Experimentation / Heart Valves / Hematocrit Limits: Animals Language: Korean Journal: The Korean Journal of Thoracic and Cardiovascular Surgery Year: 1997 Type: Article