Your browser doesn't support javascript.
loading
Accuracy, Precision, and Validity of Fever Detection using Non-invasive Temperature Measurement in Adult Coronary Care Unit Patients with Pulmonary Catheters
Journal of Korean Academy of Nursing ; : 424-433, 2012.
Article in Korean | WPRIM | ID: wpr-200560
ABSTRACT

PURPOSE:

To investigate the accuracy, precision and validity of fever detection of tympanic membrane (TM), temporal artery (TA) and axillary temperature (AT) compared with pulmonary artery temperature (PA).

METHODS:

Repeated-measures design was conducted for one year on 83 adult cardiac care unit patients with pulmonary artery catheters after open heart surgery. Sequential temperature measurements were taken three times at 20-minute intervals. Accuracy, precision, repeatability, and validity of fever detection were analyzed.

RESULTS:

Mean pulmonary artery temperature was 37.04degrees C (SD 0.70degrees C). The mean (SD) offsets from PA, with the mean reflecting accuracy and SD reflecting precision, were -1.31degrees C (0.75degrees C) for TA, -0.20degrees C (0.24degrees C) for TM, and -0.97degrees C (0.64degrees C) for AT. Percentage of pairs with differences within +/-0.5degrees C was 9.6% for TA, 19.7% for AT, and 91.6% for TM. Repeated measurements with all three methods had mean SD values within 0.04degrees C. Sensitivity, specificity, and positive and negative predictive values of tympanic measurements were 0.76, 1.0, and 1.0, and 0.90, respectively.

CONCLUSION:

Results show that TM best reflects PA, and is most consistent, accurate, and precise. AT tends to underestimate PA, and TA is least accurate and precise. Therefore tympanic membrane measurement is a reliable alternative to other non-invasive methods of measuring temperatures.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Pulmonary Artery / Axilla / Temporal Arteries / Thermometers / Tympanic Membrane / Body Temperature / Catheterization, Swan-Ganz / Cardiac Care Facilities / Body Mass Index / Fever Type of study: Diagnostic study / Prognostic study Limits: Aged / Female / Humans / Male Language: Korean Journal: Journal of Korean Academy of Nursing Year: 2012 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Pulmonary Artery / Axilla / Temporal Arteries / Thermometers / Tympanic Membrane / Body Temperature / Catheterization, Swan-Ganz / Cardiac Care Facilities / Body Mass Index / Fever Type of study: Diagnostic study / Prognostic study Limits: Aged / Female / Humans / Male Language: Korean Journal: Journal of Korean Academy of Nursing Year: 2012 Type: Article