The Journal of the Korean Society for Therapeutic Radiology and Oncology ; : 100-106, 2003.
Article in Korean | WPRIM | ID: wpr-200729
ABSTRACT

PURPOSE:

To develop a patients' setup verification tool (PSVT) to verify the alignment of the machine and the target isocenters, and the reproducibility of patients' setup for three dimensional conformal radiotherapy (3DCRT) and intensity modulated radiotherapy (IMRT). The utilization of this system is evaluated through phantom and patient case studies. MATERIALS AND

METHODS:

We developed and clinically tested a new method for patients' setup verification, using digitally reconstructed radiography (DRR), simulation, portal and digital images. The PSVT system was networked to a Pentium PC for the transmission of the acquired images to the PC for analysis. To verify the alignment of the machine and target isocenters, orthogonal pairs of simulation images were used as verification images. Errors in the isocenter alignment were measured by comparing the verification images with DRR of CT images. Orthogonal films were taken of all the patients once a week. These verification films were compared with the DRR were used for the treatment setup. By performing this procedure every treatment, using humanoid phantom and patient cases, the errors of localization can be analyzed, with adjustments made from the translation. The reproducibility of the patients' setup was verified using portal and digital images.

RESULTS:

The PSVT system was developed to verify the alignment of the machine and the target isocenters, and the reproducibility of the patients' setup for 3DCRT and IMRT. The results show that the localization errors are 0.8+/-0.2 mm (AP) and 1.0+/-0.3 mm (Lateral) in the cases relating to the brain and 1.1+/-0.5 mm (AP) and 1.0+/-0.6 mm (Lateral) in the cases relating to the pelvis. The reproducibility of the patients' setup was verified by visualization, using real-time image acquisition, leading to the practical utilization of our software.

CONCLUSION:

A PSVT system was developed for the verification of the alignment between machine and the target isocenters, and the reproducibility of the patients' setup in 3DCRT and IMRT. With adjustment of the completed GUI-based algorithm, and a good quality DRR image, our software may be used for clinical applications.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Pelvis / Radiotherapy / Brain / Radiography / Radiotherapy, Conformal Type of study: Diagnostic study Limits: Humans Language: Korean Journal: The Journal of the Korean Society for Therapeutic Radiology and Oncology Year: 2003 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Pelvis / Radiotherapy / Brain / Radiography / Radiotherapy, Conformal Type of study: Diagnostic study Limits: Humans Language: Korean Journal: The Journal of the Korean Society for Therapeutic Radiology and Oncology Year: 2003 Type: Article