Your browser doesn't support javascript.
loading
Comperieon of the Tono-Pen and the Goldmann Applanation Tonometry in Csreening Test
Journal of the Korean Ophthalmological Society ; : 281-286, 1991.
Article in Korean | WPRIM | ID: wpr-203082
ABSTRACT
Tono-Pen was portable, miniaturized digital electronic tonometer. We compared Tono-Pen with Goldmann applanation tonometer through the mass screen studies on 634 eyes of 317 patients to evaluate the reliability of the Tono-Pen in determining intraocular pressure(IOP). We found 60% of the Tono-Pen readings to be within +/-1.5 mmHg of the Goldmann applanation tonometry readings and 85% to be within +/-2.5 mmHg difference. The correlation coefficient between the readings of the two instruments was 0.85. The Tono-Pen tonometry corresponded well to the Goldmann applanation tonometry in the 11 to 20 mmHg interval. We concluded that Tono-Pen tonometry is very useful in screening test to measure the intraocular pressure.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Reading / Mass Screening / Intraocular Pressure / Manometry Type of study: Screening study Limits: Humans Language: Korean Journal: Journal of the Korean Ophthalmological Society Year: 1991 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Reading / Mass Screening / Intraocular Pressure / Manometry Type of study: Screening study Limits: Humans Language: Korean Journal: Journal of the Korean Ophthalmological Society Year: 1991 Type: Article