Your browser doesn't support javascript.
loading
Estimation of Radiation Exposure of 128-Slice 4D-Perfusion CT for the Assessment of Tumor Vascularity
Korean Journal of Radiology ; : 547-552, 2010.
Article in English | WPRIM | ID: wpr-207985
ABSTRACT

OBJECTIVE:

We aimed to estimate the effective dose of 4D-Perfusion-CT protocols of the lung, liver, and pelvis for the assessment of tumor vascularity. MATERIALS AND

METHODS:

An Alderson-Rando phantom equipped with thermoluminescent dosimeters was used to determine the effective dose values of 4D-Perfusion-CT. Phantom measurements were performed on a 128-slice single-source scanner in adaptive 4D-spiral-mode with bidirectional table movement and a total scan range of 69 mm over a time period of nearly 120 seconds (26 scans). Perfusion measurements were simulated for the lung, liver, and pelvis under the following conditions lung (80 kV, 60 mAs), liver (80 kV/80 mAs and 80 kV/120 mAs), pelvis (100 kV/80 mAs and 100 kV/120 mAs).

RESULTS:

Depending on gender, the evaluated body region and scan protocol, an effective whole-body dose between 2.9-12.2 mSv, was determined. The radiation exposure administered to gender-specific organs like the female breast tissue (lung perfusion) or to the ovaries (pelvic perfusion) led to an increase in the female specific dose by 86% and 100% in perfusion scans of the lung and the pelvis, respectively.

CONCLUSION:

Due to a significant radiation dose of 4D-perfusion-CT protocols, the responsible use of this new promising technique is mandatory. Gender- and organ-specific differences should be considered for indication and planning of tumor perfusion scans.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Radiation Dosage / Phantoms, Imaging / Four-Dimensional Computed Tomography / Neoplasms Limits: Female / Humans / Male Language: English Journal: Korean Journal of Radiology Year: 2010 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Radiation Dosage / Phantoms, Imaging / Four-Dimensional Computed Tomography / Neoplasms Limits: Female / Humans / Male Language: English Journal: Korean Journal of Radiology Year: 2010 Type: Article