Performance Evaluation of the Autokit Total Ketone Bodies
Journal of Laboratory Medicine and Quality Assurance
;
: 178-180, 2017.
Article
in Korean
| WPRIM
| ID: wpr-209174
ABSTRACT
In this study, the Autokit Total Ketone Bodies kit (Wako Pure Chemical, Japan), a total ketone measurement assay using an enzymatic method, was evaluated using a Roche Cobas e702 instrument (Roche Diagnostics, Germany). Precision, linearity, carryover, and reference range verification were evaluated with reference to Clinical Laboratory Standards Institute guidelines. Standard materials provided by the manufacturer and patient samples were used for the evaluation. The precision and carryover of the evaluation result satisfied the acceptance criteria. Linearity was also acceptable at more than 0.99. The quantitative Autokit Total Ketone Bodies kit is precise, and can be widely used in clinical laboratories.
Full text:
Available
Index:
WPRIM (Western Pacific)
Main subject:
Reference Values
/
3-Hydroxybutyric Acid
/
Evaluation Studies as Topic
/
Ketone Bodies
/
Methods
Type of study:
Diagnostic study
/
Evaluation studies
/
Practice guideline
Limits:
Humans
Language:
Korean
Journal:
Journal of Laboratory Medicine and Quality Assurance
Year:
2017
Type:
Article