ABSTRACT
STATEMENT OF

PROBLEM:

In order to increase preload with reducing the friction coefficient, abutment screws coated with pure gold and Teflon as dry lubricant coatings have been introduced. But the reported data indicate that if screw repeated tightening and loosening cycle, an efficiency of increasing preload was decreased by screw surface wearing off.

PURPOSE:

This study was to evaluate the influence of tungsten carbide/carbon coating, which has superior hardness and frictional wear resistance, on the preload of abutment screws and the stability of coating surface after repeated closures. MATERIAL AND

METHOD:

The rotational values of abutment screws and the compressive forces between abutment and fixture were measured in implant systems with three different joint connections, one external butt joint and two internal cones. Moreover, the stability and the alteration of coating surface were examined by comparison of the compressive force and the removable torque values during 10 consecutive trials, observation with scanning electron microscope and analyzed the elemental composition with energy dispersive x-ray spectroscopy. RESULTS AND

CONCLUSION:

1. Application of coating resulted in significant increase of compressive force in all implant systems(P.05). 3. Removable torque values were greater with non-coated screw than that with coated screw (P.05). 5. After repeated trials, the surface layer of coated screw was maintained relatively well. However, surface wearing and irregular titanium fragments were found in non-coated screw.