The influence of abutment screw tightening timing and DLC coating of conical connection implant system / 대한치과보철학회지
The Journal of Korean Academy of Prosthodontics
;
: 209-216, 2008.
Article
in Korean
| WPRIM
| ID: wpr-226230
ABSTRACT
STATEMENT OF PROBLEM:
It is difficult about assessing the loosening tendency of conical connection type implant after cyclic loading with diamond like carbon coating one-piece abutments, and also about the retightening effect.PURPOSE:
This study was performed to investigate the influence of one-piece abutment screw retightening after 5.0X10(4) cyclic loading and Diamond like coating MATERIAL ANDMETHODS:
Thirty two ITI implant were divided to 4 groups. Group 1,3-titanium abutment, group 2,4 - diamond like carbon coated abutment. Group 1,2 - 20.0X10(4) cyclic loading after 5.0X10(4) cyclic loading, Group 3,4- after 20.0X10(4) Cyclic loading. After cyclic loading, periotest values were taken and removal torque values of abutments were measured with a digital torque gauge.RESULTS:
1. The removal torque of group 2 after 5.0X10(4) cyclic loading is slightly greater than the other groups but not significantly higher than others (P>0.05). 2. The final removal torque values after 20.0X10(4) cyclic loading of group 1 is bigger than group 3, and group 2 is bigger than group 4, but not significantly higher (P>0.05). 3. The final removal torque values after 20.0X10(4) cyclic loading of all groups are not significantly different (P>0.05).