Your browser doesn't support javascript.
loading
Influence of pivot distance on the performance of a mechanic heart valve / 生物医学工程学杂志
Journal of Biomedical Engineering ; (6): 1037-1040, 2014.
Article in Zh | WPRIM | ID: wpr-234463
Responsible library: WPRO
ABSTRACT
Based on the analysis of the influence of the valve pivot distance on the performance of mechanical heart valve (MHV), such as the valve opening and closing features, flow field characteristics and the valve assembly properties, value constraints of the valve pivot distance were established, and the reasonable valve was obtained by means of the finite element method. It can be shown that the central flow characteristics of the valve could be enhanced with the increasing of the ratio of pivot distance to valve inner diameter, but the plastic deformation of the ring could be liable to occur in the MHV assembly process. It is proved that the valve of specifications can be designed in similar ratio of pivot distance to valve inner diameter according to the result of the valve performance experiment.
Subject(s)
Full text: 1 Index: WPRIM Main subject: Prosthesis Design / Heart Valve Prosthesis / Heart Valves Language: Zh Journal: Journal of Biomedical Engineering Year: 2014 Type: Article
Full text: 1 Index: WPRIM Main subject: Prosthesis Design / Heart Valve Prosthesis / Heart Valves Language: Zh Journal: Journal of Biomedical Engineering Year: 2014 Type: Article