It is necessary to monitor intraoperative spinal function in order to prevent spinal neurological deficit during spine surgery. This study aims to extract characteristic electrophysiological monitoring parameters during surgical treatment of scoliosis. The problem, "the monitoring parameters in time domain are of great variability and are sensitive to noise", may also be solved in this study. By use of continuous wavelet transform to analyze the intraoperative cortical somatosensory evoked potential (CSEP), three new characteristic monitoring parameters in time-frequency domain (TFD) are extracted. The results indicate that the variability of CSEP characteristic parameters in TFD is lower than the variability of those in time domain. Therefore, the TFD characteristic monitoring parameters are more stable and reliable parameters of latency and amplitude in time domain. The application of TFD monitoring parameters during spine surgery may avoid spinal injury effectively.