HPLC-fluorescent spectrometric determination of serum mexiletine concentration after derivatization with fluram / 药学学报
Acta Pharmaceutica Sinica
;
(12): 215-217, 2003.
Article
in Chinese
| WPRIM
| ID: wpr-251139
ABSTRACT
<p><b>AIM</b>To establish an HPLC-fluorescent spectrometric method for the determination of mexiletine hydrochloride in plasma after derivatization with fluram.</p><p><b>METHODS</b>Fluram acetone solution was added to the deproteinized plasma with acetone to obtain the derivative of mexiletine. The HPLC method was performed on a column of Allitima C18 (150 mm x 4.6 mm, 5 microns) with the mobile phase of methanol-water-diethylamine-phosphoric acid buffer (2.4 mol.L-1, pH 4.0) (70282), and the detective wavelength were set at Ex 392 nm and Em 480 nm.</p><p><b>RESULTS</b>Mexiletine has a liner range over the concentration range from 0.100-6.400 mg.L-1. The lowest detectable concentration of this method was 5 micrograms.L-1 (S/N > or = 4). The intra-day and inter-day RSDs were 1.34%-5.31%, respectively.</p><p><b>CONCLUSION</b>This method is simple, selective and can be used for therapeutic drug monitoring (TDM) and pharmacokinetic studies of mexiletine.</p>
Full text:
Available
Index:
WPRIM (Western Pacific)
Main subject:
Blood
/
Pharmacokinetics
/
Chemistry
/
Chromatography, High Pressure Liquid
/
Fluorescamine
/
Methods
/
Mexiletine
/
Anti-Arrhythmia Agents
Limits:
Humans
Language:
Chinese
Journal:
Acta Pharmaceutica Sinica
Year:
2003
Type:
Article
Similar
MEDLINE
...
LILACS
LIS