Analysis on DQA protocol of fMRI / 生物医学工程学杂志
Journal of Biomedical Engineering ; (6): 1247-1250, 2010.
Article in Chinese | WPRIM | ID: wpr-260899
ABSTRACT
Our purpose is to introduce and analyze the data quality assurance (DQA) protocol of functional magnetic resonance imaging (fMRI). A water phantom was scanned to get DQA indexes. An fMRI sequence was used to get signal noise ratio (SNR) and Drift, which was calculated from maximum difference ratio of the average signal intensity in the region of interest (ROI) of image serials. The long period application of this method demonstrated that this DQA protocol can reflect imaging performance and the state of stability of the MRI scanner. Some application experience and discussion involved in DQA were also presented here.
Subject(s)
Full text: Available Index: WPRIM (Western Pacific) Main subject: Quality Control / Reference Standards / Algorithms / Image Processing, Computer-Assisted / Artificial Intelligence / Magnetic Resonance Imaging / Artifacts / Phantoms, Imaging / Methods Limits: Humans Language: Chinese Journal: Journal of Biomedical Engineering Year: 2010 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Quality Control / Reference Standards / Algorithms / Image Processing, Computer-Assisted / Artificial Intelligence / Magnetic Resonance Imaging / Artifacts / Phantoms, Imaging / Methods Limits: Humans Language: Chinese Journal: Journal of Biomedical Engineering Year: 2010 Type: Article