Your browser doesn't support javascript.
loading
In vitro cytotoxicity test of medical ultrasonic couplant / 中国医疗器械杂志
Article in Zh | WPRIM | ID: wpr-264232
Responsible library: WPRO
ABSTRACT
The cytotoxicity of medical ultrasonic couplant was tested by MTT assay and agar overlay test. By MTT assay, when the inoculum density was high, the cytotoxicity level was low, or vice versa. The cytotoxicity grade tested by agar overlay was not accord to MTT assay's too. MTT assay is suitable to test the cytotoxicity of medical ultrasonic couplant because it is quantitative and more sensitive, however, the experimental condition and the preparative method of extraction should be determined.
Subject(s)
Full text: 1 Index: WPRIM Main subject: Ultrasonics / Cytotoxicity Tests, Immunologic / Cell Line / Colorimetry / Methods Limits: Animals Language: Zh Journal: Chinese Journal of Medical Instrumentation Year: 2013 Type: Article