FE modeling and analysis of the locking plate for proximal humerus fracture / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation
;
(6): 84-87, 2013.
Article
in Chinese
| WPRIM
| ID: wpr-264267
ABSTRACT
The FE model of proximal humeral locking plate was established and analyzed to provide academic support of the proximal humeral locking plate failure. The stress distributions of the locking plate were observed under the situations of the fracture area providing support and without support. In summary, this type of locking plate cannot afford the stress at 120 degrees abduction position under the situation of fracture area provides without support. In order to reduce the probability of locking plate failure, it is recommended that using other type of locking plate instead or avoiding 120 degree abduction of the patient.
Full text:
Available
Index:
WPRIM (Western Pacific)
Main subject:
Stress, Mechanical
/
Bone Plates
/
Materials Testing
/
Prosthesis Failure
/
Equipment Failure Analysis
/
Finite Element Analysis
/
Humeral Fractures
Limits:
Humans
Language:
Chinese
Journal:
Chinese Journal of Medical Instrumentation
Year:
2013
Type:
Article
Similar
MEDLINE
...
LILACS
LIS