The Dose Effect of Isocenter Selection during IMRT Dose Verification with the 2D Chamber Array / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 222-224, 2015.
Article in Chinese | WPRIM | ID: wpr-265658
ABSTRACT
To investigate the dose effect of isocenter difference during IMRT dose verification with the 2D chamber array. The samples collected from 10 patients were respectively designed for IMRT plans, the isocenter of which was independently defined as P(o), P(x) and P(y). P(o) was fixed on the target center and the other points shifted 8cm from the target center in the orientation of x/y. The PTW729 was used for 2D dose verification in the 3 groups which beams of plans were set to 0 degrees. The γ-analysis passing rates for the whole plan and each beam were gotten using the different standards in the 3 groups, The results showed the mean passing rate of γ-analysis was highest in the P(o) group, and the mean passing rate of the whole plan was better than that of each beam. In addition, it became worse with the increase of dose leakage between the leaves in P(y) group. Therefore, the determination of isocenter has a visible effect for IMRT dose verification of the 2D chamber array, The isocenter of the planning design should be close to the geometric center of target.
Subject(s)
Full text: Available Index: WPRIM (Western Pacific) Main subject: Radiotherapy Dosage / Radiotherapy, Intensity-Modulated / Gamma Rays / Methods Limits: Humans Language: Chinese Journal: Chinese Journal of Medical Instrumentation Year: 2015 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Radiotherapy Dosage / Radiotherapy, Intensity-Modulated / Gamma Rays / Methods Limits: Humans Language: Chinese Journal: Chinese Journal of Medical Instrumentation Year: 2015 Type: Article