Validation for access recirculation and access flow rate measurement by contrast-enhanced ultrasonography during hemodialysis / 生物医学工程学杂志
Journal of Biomedical Engineering
;
(6): 84-88, 2012.
Article
in Chinese
| WPRIM
| ID: wpr-274898
ABSTRACT
To evaluate the feasibility and accuracy of contrast-enhanced ultrasonography (CEUS) for the measurement of hemodialysis access recirculation (AR) and access flow rate (Qa), a two pump system was used to simulate access and dialyzer flow. AR and Qa under different conditions, such as reversal connection of dialysis lines and the needle orientation, were compared with each other. The value of access flow and recirculation flow were calculated based on the formulas introduced in this paper, and the correlation and consistency between true flow rate and calculated values were analyzed. The measured R correlated well with true value of flow rate (r = 0.57, P = 0.038, Qa > Qb; r = 0.95, P = 0.001, Qa < Qb). The Bland-altman test showed good agreement between the calculated value based on CEUS and true values. The CEUS can be used as a new advanced technology for AR and Qa measurement.
Full text:
Available
Index:
WPRIM (Western Pacific)
Main subject:
Regional Blood Flow
/
Therapeutics
/
Blood
/
Blood Flow Velocity
/
Computer Simulation
/
Arteriovenous Shunt, Surgical
/
Ultrasonography
/
Renal Dialysis
/
Contrast Media
/
Kidney Failure, Chronic
Type of study:
Diagnostic study
Limits:
Humans
Language:
Chinese
Journal:
Journal of Biomedical Engineering
Year:
2012
Type:
Article