A Comparison of Hertel Exophthalmometer Measurements Between Various Manufacturer`s Models
Journal of the Korean Ophthalmological Society ; : 2349-2356, 1999.
Article in Korean | WPRIM | ID: wpr-28259
ABSTRACT
To detect differences among the measurements of the Hertel exophthalmometers at different bar sizes and with different anufacturer`s models, exophthalmometries were performed for 27 normal subjects and 24 patients with thyroid ophthalmopathy at the three different bar sizes and with the models of three different manufacturers. The measurements decreased with the decrease of the bar size and increased with the increase of the bar size in all situations. The measurements of three different manufacturer`s models at the same bar size were different from each other. Therefore, Hertel exophthalmometries should be performed at the same bar size and with the same manufacturer`s model to find the exact changes of exophthalmos between repetitious examinations.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Thyroid Gland / Exophthalmos Limits: Humans Language: Korean Journal: Journal of the Korean Ophthalmological Society Year: 1999 Type: Article