Your browser doesn't support javascript.
loading
Comparison between hemolysis percentage measurement and hemiglobincyanide measurement for standardizing the evaluation of hemolytic properties of biomaterials / 生物医学工程学杂志
Journal of Biomedical Engineering ; (6): 111-137, 2004.
Article in Chinese | WPRIM | ID: wpr-291171
ABSTRACT
In this study, two methods--the supernatant hemoglobin spectrophotometry, i.e. hemolysis percentage measurement(according to ISO/TR 7405) and the hemiglobincyanide measurement(according to ISO 10993-4)--were used to assay the hemolytic properties of hydroxyapatite(bioceramics) and collagen (polymer). The results showed that the conclusions drawn from using the two methods were basically consistent, and the latter was more sensitive, stable and comparable. However, some of the procedures in the hemiglobincyanide method were not defined in details. So based on our experiments we have offered some suggestions and improvements, which do not deviate from ISO and ASTM standards, for hiher practicability of usig it in standardizing the evaluation of the hemolytic properties of biomaterials. Hemiglobincyanide measurement is worthy of wider application.
Subject(s)
Full text: Available Index: WPRIM (Western Pacific) Main subject: Materials Testing / Hemoglobins / Methemoglobin / Collagen / Durapatite / Hemolysis / Methods Type of study: Practice guideline Limits: Animals / Humans Language: Chinese Journal: Journal of Biomedical Engineering Year: 2004 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Materials Testing / Hemoglobins / Methemoglobin / Collagen / Durapatite / Hemolysis / Methods Type of study: Practice guideline Limits: Animals / Humans Language: Chinese Journal: Journal of Biomedical Engineering Year: 2004 Type: Article