Your browser doesn't support javascript.
loading
Design of microvolt T wave alternans detecting system based on wireless and USB communication / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 419-423, 2008.
Article in Chinese | WPRIM | ID: wpr-309561
ABSTRACT
In this paper, a T wave alternans detection system based on wireless and USB communication is presented. The ECG signal is amplified in the first step. The amplifier has a gain of 2000 and the CMRR is 80dB. Then the signal is sampled with a 12-bit AD converter. At last, the sample result is transmitted to the PC via wireless communication and USB bus. In the PC, the T wave alternans are detected using the spectral method. Results show that this system has high data transmission rate and real time data processing capability, and can realize the detection of T wave alternans.
Subject(s)
Full text: Available Index: WPRIM (Western Pacific) Main subject: Arrhythmias, Cardiac / Telemetry / Algorithms / Computer Communication Networks / Signal Processing, Computer-Assisted / Artificial Intelligence / Diagnosis / Electrocardiography / Equipment Design / Methods Type of study: Diagnostic study Limits: Humans Language: Chinese Journal: Chinese Journal of Medical Instrumentation Year: 2008 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Arrhythmias, Cardiac / Telemetry / Algorithms / Computer Communication Networks / Signal Processing, Computer-Assisted / Artificial Intelligence / Diagnosis / Electrocardiography / Equipment Design / Methods Type of study: Diagnostic study Limits: Humans Language: Chinese Journal: Chinese Journal of Medical Instrumentation Year: 2008 Type: Article