Liquid gastric-emptying measurement using an electrical bio-impedance method / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 253-256, 2008.
Article in Chinese | WPRIM | ID: wpr-309602
ABSTRACT
This paper introduces a 4-electrode gastric-emptying measurement system using an electrical bio-impedance method by which gastric-emptying experiments of 24 healthy volunteers have been carried out. The test results show that the base impedance is stable, the average GET/2 is 8.78 +/- 1.76 min. The new system provides a new non-invasive measurement method for the clinical study on gastric-emptying functions.
Subject(s)
Full text: Available Index: WPRIM (Western Pacific) Main subject: Physiology / Electric Impedance / Eating / Gastric Emptying Limits: Humans Language: Chinese Journal: Chinese Journal of Medical Instrumentation Year: 2008 Type: Article