Study on cytotoxicity tests of medical devices based on IC50 / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 433-467, 2014.
Article in Chinese | WPRIM | ID: wpr-310303
ABSTRACT
To discuss IC50 application in cytotoxicity tests of medical devices, we firstly investigated the vibrating condition and endpoint of MTT method specified in ISO 10993-5 2009. Furthermore, we demonstrated the application of IC50 in the result evaluation of MTT method. The experimental results show that usage of IC50 in quantitative evaluation of MTT method is feasible.
Subject(s)
Full text: Available Index: WPRIM (Western Pacific) Main subject: Reference Standards / Toxicity Tests / Inhibitory Concentration 50 / Equipment and Supplies / Methods Language: Chinese Journal: Chinese Journal of Medical Instrumentation Year: 2014 Type: Article