Your browser doesn't support javascript.
loading
A project of magnetic coils newly designed to restrain the negative value of the intensity of magnetic induced electric field / 生物医学工程学杂志
Journal of Biomedical Engineering ; (6): 45-59, 2003.
Article in Chinese | WPRIM | ID: wpr-311112
ABSTRACT
In the non-invasive magnetic nerve stimulation of human brain, the coil of figure "8" is mostly used. The peak maximum is corresponding to the center of the coil, and the negative peak value is corresponding to the twain edges of the coil, the latter one is apt to cause abnormal nerve excitation. In this text, we put emphasis on the study of abnormal stimulation induced in magnetic stimulation and analyse the electric field intensity of the figure-8 coil. Aiming at the large negative value of intensity, we develop a new structure coil to restrain abnormal stimulation. Theoretical simulation is demonstrated here also.
Subject(s)
Full text: Available Index: WPRIM (Western Pacific) Main subject: Computer Simulation / Electric Stimulation / Electromagnetic Fields / Equipment Design / Electromagnetic Phenomena / Methods Language: Chinese Journal: Journal of Biomedical Engineering Year: 2003 Type: Article