Study on a non-invasive wide-range blood oxygen saturation detecting method and its instrument design / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 23-25, 2008.
Article in Chinese | WPRIM | ID: wpr-323230
ABSTRACT
This paper proposes a nonlinear method of detecting the blood oxygen saturation (SpO2) in a wide range on the basis of spectrophotometry, solves the nonlinear problem in detecting SpO2 in a wide range, and improves the applicable range and detection precision. CAN bus communication and DSP fast digital signal processing are utilized in the design of a non-invasive pulse blood oxygen saturation (SpO2) monitor in a wide range, which has a measure error below 3% within a range of 35% approximately 100%, and is applicable for clinical monitoring and detection of cosmonauts' physiological signals. This oximeter's components and their functions are also introduced in detail in the paper.
Subject(s)
Full text: Available Index: WPRIM (Western Pacific) Main subject: Oximetry / Equipment Design / Methods Language: Chinese Journal: Chinese Journal of Medical Instrumentation Year: 2008 Type: Article