Your browser doesn't support javascript.
loading
A new shockable rhythm detection algorithm for AED / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 320-322, 2009.
Article in Chinese | WPRIM | ID: wpr-329315
ABSTRACT
EMG, This paper introduces a slope variability algorithm which uses the CU Database and the AHA Database. The sensitivity, specificity, positive predictivity and accuracy are calculated and these values are compared with the results from two earlier detection algorithms. The results show this algorithm is more effective and fast than other algorithms.
Subject(s)
Full text: Available Index: WPRIM (Western Pacific) Main subject: Algorithms / Electric Countershock / Predictive Value of Tests / Sensitivity and Specificity / Defibrillators Type of study: Diagnostic study / Prognostic study Language: Chinese Journal: Chinese Journal of Medical Instrumentation Year: 2009 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Algorithms / Electric Countershock / Predictive Value of Tests / Sensitivity and Specificity / Defibrillators Type of study: Diagnostic study / Prognostic study Language: Chinese Journal: Chinese Journal of Medical Instrumentation Year: 2009 Type: Article