Evaluation of the performance of the electrochemiluminescent immunoassay system of Elecsys 2010 / 生物医学工程学杂志
Journal of Biomedical Engineering
;
(6): 673-675, 2002.
Article
in Chinese
| WPRIM
| ID: wpr-340939
ABSTRACT
Performance of the electrochemiluminescent immunoassay system of Elecsys 2010 was evaluated by quantitative assay. The functional sensitivity of the TSH assay was 0.004 microU/ml. The extra-assay imprecision was RSD = 1.45%-4.65% in the range of TSH concentrations of 0.43-88.15 microU/ml. The intra-assay imprecision was RSD = 1.38%-3.14% in the range of TSH concentrations of 0.41-87.28 microU/ml. The recovery of TSH was in the range of 98.6%-104.4%. The correlation between theoretical values and obtained values of TSH was adequate(r = 0.999). We conclude that the Elecsys 2010 is provably satisfactory in sensitivity, precision, accuracy, reliability, practicability and rapidity.
Full text:
Available
Index:
WPRIM (Western Pacific)
Main subject:
Reference Standards
/
Blood
/
Immunoassay
/
Thyrotropin
/
Sensitivity and Specificity
/
Electrochemistry
/
Evaluation Studies as Topic
/
Luminescent Measurements
Type of study:
Diagnostic study
/
Evaluation studies
Limits:
Humans
Language:
Chinese
Journal:
Journal of Biomedical Engineering
Year:
2002
Type:
Article
Similar
MEDLINE
...
LILACS
LIS