Analysis residues of perfluorooctane used in ophthalmic operation / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation
;
(6): 373-374, 2012.
Article
in Chinese
| WPRIM
| ID: wpr-342924
ABSTRACT
Analyzed the purity of the Perfluorooctane used for Ophthalmic operation using the FID (Flame Ionization Detector) method after the qualitative analysis of that. Research and discussion of the 1--hydrogen perfluorocarbon octane, which is the typical impurity of Perfluorooctane. The standard recovery of the 1--hydrogen perfluorocarbon octane is 93%-97% within the range of linearity, and the RSD (relative standard deviation) is below 2%. The above test method has the advantages of easy operation, stability and accuracy. So it is a good reference support to bring the specification of quality control of the Perfluorooctane which is used for Ophthalmic operation into YY of medical device.
Full text:
Available
Index:
WPRIM (Western Pacific)
Main subject:
Ophthalmology
/
Quality Control
/
Chemistry
/
Fluorocarbons
/
Methods
Type of study:
Qualitative research
Language:
Chinese
Journal:
Chinese Journal of Medical Instrumentation
Year:
2012