Your browser doesn't support javascript.
loading
Evaluation of the levels of auto-PEEP in the patients with acute respiratory failure required mechanical ventilation
Journal of Medical Research ; : 46-51, 2004.
Article in Vietnamese | WPRIM | ID: wpr-3607
ABSTRACT
We have studied 27 patients on medical ventilation in whom we measured auto-PEEP by means of end expiratory airway occlusion. Four patients with ARDS had 10.25  2.05 cm H2O of auto PEEP. Six patients with COPD had 12.83  4.98 cm H2O of auto PEEP. Seven patients with acute severe asthma had 20.29  7.21 cm H2O of auto PEEP. Ten patients with neuromuscular disease had 5.2  4.07 cm H2O of auto PEEP. PEEP was set equal to half the auto-PEEP to treat patients with ARDS, COPD and acute severe asthma. We have show that the value of auto PEEP fell to 4.67  3.7 cm H2O in patient with ARDS, to 7  2.31 in patient of COPD and to 9.71  1.67 in patients of asthma. PEEP 50% of auto-PEEP is safe and does not worsen pulmonary hyperinflation
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Respiration, Artificial / Respiratory Insufficiency / Positive-Pressure Respiration, Intrinsic Language: Vietnamese Journal: Journal of Medical Research Year: 2004 Type: Article